厦门机电-关于招标文件XM2015-DZ0518母婴动态监护系统(Holter)采购更改通知
致XM2015- DZ0518项目投标人:
接采购人通知,现对XM2015- DZ0518母婴动态监护系统(Holter)采购招标文件“二、主要技术性能指标及要求、配置:”做如下更改:
*19、配有专业智能自动分析系统,可分析以下内容
(1)Time (HH:MM)事件出现时间
(2)Criteria (Met/Not Met)决策支持(匹配/不匹配)
(3)Loss %信号丢失率与成功率
(4)Mean (bpm)胎儿心率均值
(5)Basal (bpm)基线
(6)Small Acceleration Count (10bpm for 15sec)小加速
(7)Large Acceleration Co XM2015-DZ0518unt (15bpm for 15sec)大加速
(8)Small Deceleration Count (<20 lost beats)小减速
(9)Large Deceleration Count (>=20 lost beats)大减速
(10)STV (ms) Short Term Variation短期变异
(11)MMR (ms) Mean Minute Range, LTV平均每分种变异范围
(12)High Variation (%) (5 in 6 min > 32ms)高变异率
(13)High Variation (ms)高变异
(14)Low Variation (%) (5 in 6 min < 30ms)低变异率
(15)Low Variation (ms)低变异
(16)系统报警时间,解除报警时间。
修改成
19、配有专业智能自动分析系统,可分析以下内容(1)Time (HH:MM)事件出现时间
(2)Criteria (Met/Not Met)决策支持(匹配/不匹配)
(3)Loss %信号丢失率与成功率
(4)Mean (bpm)胎儿心率均值
(5)Basal (bpm)基线
(6)Small Acceleration Count (10bpm for 15sec)小加速
(7)Large Acceleration Count (15bpm for 15sec)大加速
(8)Small Deceleration Count (<20 lost beats)小减速
(9)Large Deceleration Count (>=20 lost beats)大减速
(10)STV (ms) Short Term Variation短期变异
(11)MMR (ms) Mean Minute Range, LTV平均每分种变异范围
(12)High Variation (%) (5 in 6 min > 32ms)高变异率
(13)High Variation (ms)高变异
(14)Low Variation (%) (5 in 6 min < 30ms)低变异率
(15)Low Variation (ms)低变异
(16)系统报警时间,解除报警时间。
技术规格及商务响应表中相应修改该条款。
本更改通知条款如与招标文件中相应条款有冲突的地方请以本更改通知为准。其他内容如在本通知中未体现,则仍以招标文件中的要求为准。请投标人接到通知后将回执回传至我司,传真为0592-2298138。否则视为贵司已经收悉。
厦门经发机电设备招标有限公司
二0一五年十月三十日
_______________________________________
回 执
厦门经发机电设备招标有限公司:
我司已收悉上述内容,现给予回传确认。(FAX:0592-2298138)
单位盖章
年 月 日